Remote control commands
R&S
®
RTM3000
515User Manual 1335.9090.02 ─ 09
MEASurement<m>:AON
Starts the quick measurement.
Usage:  Event
MEASurement<m>:AOFF
Stops the quick measurement.
Usage:  Event
MEASurement<m>:ARESult?
Returns the results of the quick measurement.
Return values: 
<Data> List of values
Quick measurement results are listed in the following order:
PEAK (Vpp), UPE (Vp+), LPE (Vp-), CYCR (RMS-Cyc), CYCM
(MeanCyc), PER (T), FREQ (f), RTIM (tr), FTIM (tf).
Usage:  Query only
MEASurement<m>:ALL[:STATe]
Starts or stops the quick measurement and sets the status bit.
Suffix: 
<m>
.
1..8
The suffix is irrelevant.
Parameters:
<State> ON | OFF
*RST:  OFF
17.7.2  Automatic measurements
● Measurement settings...........................................................................................515
● Measurements results...........................................................................................521
● Measurement gate................................................................................................ 526
● Reference levels................................................................................................... 528
17.7.2.1  Measurement settings
MEASurement<m>[:ENABle]...........................................................................................516
MEASurement<m>:MAIN................................................................................................516
MEASurement<m>:SOURce...........................................................................................518
MEASurement<m>:DELay:SLOPe...................................................................................518
Measurements