Measurements
R&S
®
RTM3000
153User Manual 1335.9090.02 ─ 09
Meas. type Symbol Description Graphic / formula
Mean Cycle MeanCyc
in V
Mean value of the left-most signal period.
DC
N T
RMS Cycle RMS-Cyc
in V
RMS (root mean square) value of the voltage of the left-
most signal period.
DC
AC
N T
Peak Peak Vpp Difference of maximum and minimum values.
Max
Min
Pk-Pk
Peak+ Vp+ Maximum value within the displayed waveform.
Max
Peak- Vp- Minimum value within the displayed waveform.
Min
Pos. Overshoot
Neg. Overshoot
+Ovr
-Ovr
in %
Overshoot of a square wave after a rising or falling edge. It
is calculated from measurement values top level, base level,
local maximum, local minimum, and amplitude.
%100
Amplitude
TopMax
Over
local
%100
Amplitude
Min Base
Over
local
Top
Base
Amplitude
Over+
Over-
Mean Value Mean Arithmetic average of the complete displayed waveform.
N
k
k
x
N
1
1
Mean
RMS Value RMS RMS (root mean square) value of the voltage of the com-
plete displayed waveform.
N
k
k
x
N
1
2
1
RMS
σ-Std. Deviation
σ
Standard deviation of the displayed waveform.
N
k
k
x
N
1
2
Mean
1
1
σ
Automatic measurements