Measurements
R&S
®
RTM3000
160User Manual 1335.9090.02 ─ 09
Figure 8-5: Cursor measurement with vertical and horizontal cursors and Set To Trace
Results = below the grid
Cursor lines 1, 2, 3 = no focus
Cursor line 4 = has focus, can be moved by turning the [Navigation] knob
To configure cursor measurements
1. Press the [Cursor] key.
The cursors are activated with the latest setting.
2. Press the [Cursor] key again.
The "Cursor" menu opens.
3. Select the cursor "Type".
4. Select the "Source", the waveform you want to measure.
5. If you want to measure different sources:
a) Enable "Second Source".
b) Select the second waveform.
6. Set additional settings if necessary: Track Scaling, Coupling, or Set To Trace.
7. To change the position of a cursor line, you can use several methods:
Drag the cursor line on the screen.
Press the [Navigation] knob repeatedly until the required cursor line is active
(marked with a solid line).
Turn the knob to move the line.
Tap the corresponding result value in the result line at the bottom.
Cursor measurements