Command Reference
R&S
®
ZNB/ZNBT
1181User Manual 1173.9163.02 ─ 54
BWAVe calibration of received wave b
1
, b
2
, .... The port num-
ber <CalPort>, the used <SourceType>, and the
<SourcePort> number must be specified in addition.
Alternative: Parameters B1 and B2.
B1 | B2 ... direct wave and cal port setting for received waves
b
1
, b
2
... The parameters B1 and B2 are ZVR-compatible. No
additional parameters need to be specified. The source for B1 is
Port 2 and vice versa.
<CalPort> Calibrated port number
Range: 1 to port number of the analyzer
<SourceType> PORT | GENerator
PORT - internal source at port <CalPort>, to be fed to port
<SourcePort> using an external through connection. If
<CalPort> = <SourcePort>, an Open or Short standard is
required.
GENerator - external generator no. <source_no>.
<SourcePort> Number of the port for the internal source or generator.
<CalOnlyPortFreq>
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is 1).
CALC:PAR:MEAS 'TRC1', 'A1'
Select a
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no. <Mk> in the center of the sweep range and
query the measurement value. The calibrated power of the refer-
ence wave a
1
is approx. equal to the default source power value.
See also example for [SENSe<Ch>:]CORRection:
OFFSet<PhyPt>[:STATe].
Usage:
Setting only
Manual operation: See "Port Overview" on page 446
[SENSe<Ch>:]CORRection:POWer<PhyPt>:AWAVe[:STATe] <Boolean>
Enables or disables the receiver power calibration for channel <Ch> and for the refer-
ence waves a
n
. The setting command is disabled unless the reference waves have
been power calibrated ([SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
on page 1180 AWAVe,...). The query always returns a result.
SCPI Command Reference