Command Reference
R&S
®
ZNB/ZNBT
1182User Manual 1173.9163.02 ─ 54
This command extends the functionality of manual control, where the reference
receiver calibration is included in the source power calibration of each port (and always
ON).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the receiver power calibration
for the reference waves a
n
.
*RST: OFF
Example: See [SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
[<CalOnlyPortFreq>]
Starts the receiver calibration (3
rd
power calibration step for intermodulation measure-
ments), stores and applies the calibration data.
Note: The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire is enabled
only after steps 1 and 2 have been carried out (SOURce<Ch>:POWer<PhyPt>:
CORRection:IMODulation:LTONe[:ACQuire], SOURce<Ch>:POWer<PhyPt>:
CORRection:IMODulation:UTONe[:ACQuire]).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Port number. This suffix is ignored; the analyzer calibrates the
port that is selected as a receive port for the intermodulation
measurement
([SENSe<Ch>:]FREQuency:IMODulation:RECeiver).
Setting parameters:
<CalOnlyPortFreq>
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
IMODulation:LTONe[:ACQuire]
Usage: Setting only
Manual operation: See "Start Cal Sweep" on page 557
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
Queries the stimulus values of the active calibration. A calibration must be selected