Command Reference
R&S
®
ZNB/ZNBT
1183User Manual 1173.9163.02 ─ 54
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt> Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S ZNB/ZNBT-K4).
Example:
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 fre-
quency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the
active channel.
Usage:
Query only
Manual operation: See "Apply" on page 439
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:ALL:COUNt?
Returns the total number of port assignments of all calibrations.
See Chapter 5.5.5.4, "Multiple Port Assignments", on page 173 for background infor-
mation.
Note:
Use [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt? to get the
number of port assignments of the "current" calibration.
Suffix:
<Ch>
.
Channel number
Example: See Chapter 9.2.5.2, "MultiCal (with Calibration Unit)",
on page 1407
Usage: Query only
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
Returns the number of port assignments of the "current" calibration, i.e. the calibration
last created using [SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure.
This is particulary useful if the default port assignments were established using
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault.
See Chapter 5.5.5.4, "Multiple Port Assignments", on page 173 for background infor-
mation.
Note:
SCPI Command Reference

Table of Contents