Manual operation: See "Edit Test Port Connection" on page 739
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:CONFigure:MTESt
{<MatrixTestPort>, <TestPort>}
Sets/gets the matrix test port connections as a comma-separated list of port numbers
Note: a manual RF connection configuration must be performed within a transaction,
enclosed by the commands SYSTem:COMMunicate:RDEVice:SMATrix:
CONFigure:STARt and SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
END. The modified configuration will not be applied until the end of this transaction.
Suffix:
<Matr>
.
Index of the switch matrix (see SYSTem:COMMunicate:
RDEVice:SMATrix<Matr>:DEFine on page 1321)
Setting parameters:
{<MatrixTestPort> Number of the switch matrix test port
<TestPort>} Number of the test port.
Test ports must be numbered subsequently starting at 1.
Example: See SYSTem:COMMunicate:RDEVice:SMATrix:
CONFigure:STARt
Manual operation: See "Edit Test Port Connection" on page 739
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:CONFigure:MVNA
{<MatrixVNAPort>, <VNAPort>}
Sets/gets the RF connections between switch matrix and VNA as a comma-separated
list of port numbers.
Note: a manual RF connection configuration must be performed within a transaction,
enclosed by the commands SYSTem:COMMunicate:RDEVice:SMATrix:
CONFigure:STARt and SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
END. The modified configuration will not be applied until the end of this transaction.
Suffix:
<Matr>
.
Configuration index of the switch matrix (see SYSTem:
COMMunicate:RDEVice:SMATrix<Matr>:DEFine
on page 1321)
Setting parameters:
{<MatrixVNAPort> Number of the switch matrix VNA port
Range: 1 ... # of VNA ports of the switch matrix
<VNAPort>} Number of the VNA port
Range: 1 ... # of RF ports of the VNA
Example: See SYSTem:COMMunicate:RDEVice:SMATrix:
CONFigure:STARt
Manual operation: See "Edit Matrix VNA Port Connections" on page 738