Manual operation: See "Edit Test Port Connection" on page 739
SYSTem:COMMunicate:RDEVice:SMATrix:COUNt?
Gets the number of configured switch matrices (see SYSTem:COMMunicate:
RDEVice:SMATrix<Matr>:DEFine on page 1321)
Usage: Query only
Manual operation: See "Configured Devices" on page 733
SYSTem:COMMunicate:RDEVice:SMATrix:DELete
Unregisters all switch matrices, i.e. removes them from the list of configured devices.
Example:
See SYSTem:COMMunicate:RDEVice:SMATrix:
CONFigure:STARt
Usage: Event
Manual operation: See "Delete All" on page 735
SYSTem:COMMunicate:RDEVice:SMATrix:SCAN?
Scans for external switch matrices connected via USB or Direct Control.
Returns a comma separated list
<Driver_1>,<Interface_1>,<Address_1>,...,<Driver_N>,
<Interface_N>,<Address_N>, one triple
<Driver_n>,<Interface_n>,<Address_n> for each detected switch matrix
<Matr>=1,...,N.
Use SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:DEFine to configure them.
Usage:
Query only
Manual operation: See "Scan Instruments" on page 733
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:DEFine <Unused>, <Driver>,
<Interface>, <Address>
This command allows
to configure (i.e. register) a new switching matrix
to modify the (management) connection settings of an already configured switching
matrix
to get the (management) connection settings of a configured switching matrix
Suffix:
<Matr>
.
Index of the switch matrix.
Switch matrices are numbered in ascending order, starting with
1. If a number is re-used, the current configuration is overwritten.