Command Reference
R&S
®
ZNB/ZNBT
1353User Manual 1173.9163.02 ─ 54
Example:
TRIGger:CHANnel:AUXiliary:ENABle ON
Turn on the trigger output for channel no. 1 (channel suffix omit-
ted).
TRIGger:CHANnel:AUXiliary:DURation 10 us
Change the width of the output trigger pulses.
TRIGger:CHANnel:AUXiliary:POSition BEFore
Send the output trigger pulses before data acquisition starts.
TRIGger:CHANnel:AUXiliary:INTerval SWEep
Send one output trigger pulse per sweep.
TRIGger:CHANnel:AUXiliary:OPOLarity POSitive
Select positive polarity for the output trigger pulses.
TRIGger:CHANnel<Ch>:AUXiliary<n>:DURation <TrigOutDuration>
Specifies the width of the output trigger pulses at EXT TRIG OUT.
The trigger duration must be shorter than the selected sweep interval (see TRIGger:
CHANnel<Ch>:AUXiliary<n>:INTerval).
If the trigger duration is too long, you can modify one of the following settings:
Reduce the trigger duration, if possible.
Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<TrigOutDuration> Pulse width between 1 μs (1E-6 s) and 1 s.
*RST: 1 us
Default unit: s
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
on page 1352
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval <Type>
Specifies whether the trigger output pulses are sent once every measurement point or
once every sweep.
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
SCPI Command Reference