Command Reference
R&S
®
ZNB/ZNBT
856User Manual 1173.9163.02 ─ 54
Example:
*RST; CALCulate:LIMit:CIRCle:DATA 0, 0, 0.5
Define a circle limit line centered around the origin of the polar
diagram, assigning a radius of 0.5 U.
CALCulate:FORMat POLar
CALCulate:LIMit:CIRCle:DISPlay ON
Activate a polar diagram and show the circle limit line in the dia-
gram.
Manual operation: See "Show Limit Circle" on page 356
CALCulate<Chn>:LIMit:CIRCle:FAIL?
Returns a 0 or 1 to indicate whether or not the circle limit check has failed. 0 repre-
sents pass and 1 represents fail
Tip: Use CALCulate:CLIMits:FAIL? to perform a composite (global) limit check.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See CALCulate<Chn>:LIMit:CIRCle[:STATe]
Usage: Query only
Manual operation: See "Limit Check" on page 357
CALCulate<Chn>:LIMit:CIRCle:SOUNd[:STATe] <Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded circle limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Fail beep on or off.
*RST: OFF
Example:
CALCulate:LIMit:CIRCle:STATe ON; SOUNd ON
Switch the limit check on and activate the fail beep.
Manual operation: See "Limit Fail Beep" on page 358
CALCulate<Chn>:LIMit:CLEar
Resets the limit check results for the limit line test.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage: Event
Manual operation: See "Clear Test" on page 345