Remove the cover
1. Unscrew the retaining ring shown on the cover and lift the
transparent cover off of the housing.
Position the decimal point and select the meter function
2. Press the left and right configuration buttons
simultaneously and release them immediately.
3. To move the decimal point to the desired location, press
the left configuration button. Note that the decimal point
wraps around.
4. To scroll through the mode options, press the right
configuration button repeatedly until the desired mode
is displayed. See Table 1.
Store the information
5. Press both configuration buttons simultaneously for two
seconds. Note that the meter displays “----” for
approximately 7.5 seconds while the information is being
stored.
Set the display equivalent to a 4 mA signal
6. Press the left button for two seconds.
7. To decrease the display numbers, press the left
configuration button and to increase the numbers,
press the right configuration button. Set the numbers
between –999 and 1000.
8. To store the information, press both configuration buttons
simultaneously for two seconds. Note that the meter
displays “----” for approximately 7.5 seconds while the
information is being stored.
Set the display equivalent to a 20 mA signal
9. Press the right button for two seconds.
10. To decrease the display numbers, press the left
configuration button on the display and to increase the