LCD METER SPECIFICATIONS
Input Signal
4–20 mA dc.
Meter Indication
4-digit LCD showing –999 to 9999. A 20-segment bar graph
directly represents the 4–20 mA current.
Digital Display Resolution
0.05% of range ±1 digit.
Analog Bar Graph Resolution
1 mA or 5% of input range.
Indication Accuracy
0.25% of range ±1 digit.
Scaling/Configuration
4 mA Point Limits: –999 to 1000.
Span limits: 0200 to 9999.
The sum of the 4 mA point and span must not exceed 9999.
Adjustments are made using non-interactive zero and span
buttons.
Temperature Limits
Storage: –40 to 85 °C (–40 to 185 °F).
Operating: –20 to 70 °C (–4 to 158 °F).
Between temperatures –40 to –20 °C (–40 to –4 °F),
the loop is intact and the meter is not damaged.
Temperature Effect
0.01% of range per °C on zero.
0.02% of range per °C on span