EasyManua.ls Logo

Rosemount 1151 - LCD Meter Specifications; Input Signal and Meter Indication; Resolution and Accuracy; Scaling and Configuration Limits

Rosemount 1151
9 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCD METER SPECIFICATIONS
Input Signal
4–20 mA dc.
Meter Indication
4-digit LCD showing –999 to 9999. A 20-segment bar graph
directly represents the 4–20 mA current.
Digital Display Resolution
0.05% of range ±1 digit.
Analog Bar Graph Resolution
1 mA or 5% of input range.
Indication Accuracy
0.25% of range ±1 digit.
Scaling/Configuration
4 mA Point Limits: –999 to 1000.
Span limits: 0200 to 9999.
The sum of the 4 mA point and span must not exceed 9999.
Adjustments are made using non-interactive zero and span
buttons.
Temperature Limits
Storage: –40 to 85 °C (–40 to 185 °F).
Operating: –20 to 70 °C (–4 to 158 °F).
Between temperatures –40 to –20 °C (–40 to –4 °F),
the loop is intact and the meter is not damaged.
Temperature Effect
0.01% of range per °C on zero.
0.02% of range per °C on span

Related product manuals