Table 2
GuidanceandManufacture’sDeclaration–ElectromagneticImmunity
Thisequipmentisintendedforuseintheelectromagneticenvironmentspecied
below. User should assure that it is used in such an environment.
Immunity test IEC60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
±8kV contact
±2kV, ±4kV,
±8kV, ±15kV air
±8kV contact
±2kV, ±4kV,
±8kV, ±15kV
air
Floors should be wood, concrete
or ceramic tile. Humidity should
be at least 30% if it is synthetic
materials.
Electrical fast
transients/
bursts (EFT)
IEC 61000-
4-4
±2kV
100kHz
repetition
frequency
±2kV
100kHz
repetition
frequency
Main power quality should be
that of a typical commercial or
hospital environment.
Surges
IEC 61000-
4-5
±0.5kV, ±1kV
line-to-line
±0.5kV, ±1kV,
±2kV line-to-
ground
±0.5kV, ±1kV
line-to-line
±0.5kV, ±1kV,
±2kV line-to-
ground
10