SaeboStim Pro
Table 2 (Continued)
Voltage dips
IEC 61000-4-11
0%UT;0.5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0%UT;0.5cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Mains power quality should
be typical commercial or
hospital environment. UPS
power is recommended if
this device needs to be used
continuously.
0%UT;1cycle
and
70%UT;25/30
cycles
Single phase:
at 0°
0%UT;1cycle
and
70%UT;25/30
cycles
Single phase: at 0°
Voltage
interruptions
IEC 61000-4-11
0%UT;250/300
cycle
0%UT;
250/300 cycle
RATED power
frequency
magneticelds
IEC 61000-4-8
30A/m
50Hz or 60Hz
30A/m
50Hz or 60Hz
Power frequency magnetic
eldsshouldbeatlevels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.