SaeboStim Pro
Table 3
GuidanceandManufacture’sDeclaration–ElectromagneticImmunity
Thisequipmentshouldbeusedintheelectromagneticenvironmentspeciedbelow.
User should assure that it is used in such an environment.
Immunity
test
IEC60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF
IEC 61000-
4-6
3Vrms
150 kHz to 80
MHz
3Vrms Portable and mobile RF
communications equipment should
be used no closer to any parts than
the recommended separation distance
that calculated from the equation
applicable to the frequency of the
transmitter.
6Vrms
in ISM and
amateur radio
bands between
150 kHz and 80
MHz (a)
6Vrms Recommended separation distance:
Pd 2.1=
150 kHz to 80 MHz
Pd 2.1=
80MHz to 800 MHz
Pd 3.2=
800MHz to 2.7GHz
Radiated RF
IEC 61000-
4-3
10 V/m
80MHz to
2.7GHz
10 V/m
EPd /6=
at RF wireless communications
equipment bands (Portable RF
communications equipment
12