Table 3 (Continued)
a)TheISM(industrial,scienticandmedical)bandsbetween0.15MHzand80MHzare
6.765MHzto6.795MHz;13.553MHzto13.567MHz;26.957MHzto27.283MHz;
and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80
MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3
MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz,21.0
MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to
54.0 MHz.
b)Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular
/ cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
theelectromagneticenvironmentduetoxedRFtransmitters,anelectromagnetic
sitesurveyshouldbeconsidered.Ifthemeasuredeldstrengthinthelocationin
which this is used exceeds the applicable RF compliance level above, this should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating.
c) Field strengths should be less than 3V/m in the frequency range of 150k-80MHz.