SaeboStim Pro
Table 4 (Continued)
2 450
2 400 -
2 570
Bluetooth,
WLAN,
802.11 b/g/n.
LTE Band 7
Pulse
modulation
b)
217 Hz
2 0.3 28
5 240
5 100 -
5 800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2 0.3 95 500
5 785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m test
distance to permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz be used because while
it does not represent actual modulation, it would be worst case.
Table 5
Recommended Separation Distance between Portable and Mobile RF
Communications Equipment and the Nerve and Muscle Stimulator
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the Nerve and Muscle
Stimulator as recommended below, according to the maximum output power of the
communications equipment.