SaeboStim Pro
Table 5 (Continued)
This device can be used under the environment that radiated RF disturbances are
controlled. User should maintain a minimum distance between portable and mobile
RF communications equipment to prevent electromagnetic interference. Following
recommended distance is calculated according to the maximum output power of the
communication equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150kHz-80MHz
80MHz -800MHz
800MHz -2.7GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1
1.2 1.2 2.3
10
3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance “d” in meters can be estimated using the equation applicable to the
frequency of transmitter, where “P” is the maximum output power rating of the transmitter
in watts according to the transmitter manufacturer.
Note1: At 80M and 800MHz, the separation distance for the higher frequency range
applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is
aectedbyabsorptionandrefectionfromstructures,objectsandpeople.
17