18
Optical Materials
Code
Batch
Description Unit Size
ERM
-
FB012
a
IR Wavelength Standard
This wavelength standard is intended for use in the verification and calibration of
infra-red spectrometers within the range of spectral bandwidths quoted.
Certified Values
True Peak Position Shifts of Wavelength by Resolution
Spectral Bandwidth 0.5 cm
-1
Spectral Bandwidth 2 cm
-1
Spectral Bandwidth 4 cm
-1
3026.0 ± 0.4 cm
-1
0.00 ± 0.1 cm
-1
0.05 ± 0.1 cm
-1
1601.1 ± 0.4 cm
-1
0.05 ± 0.1 cm
-1
0.1 ± 0.1 cm
-1
1028.8 ± 0.4 cm
-1
0.1 ± 0.1 cm
-1
0.1 ± 0.1 cm
-1
698.0 ± 0.4 cm
-1
0.1 ± 0.1 cm
-1
0.3 ± 0.1 cm
-1
1 mL
ERM
-
FB020
a
UV
-
Visible Wavelength Standard for HPLC Detectors
This material consists of a high purity solution of holmium and neodymium oxide in
dilute perchloric acid.
This material is intended as a standard for use in the calibration of the wavelength
scale of ultraviolet-visible HPLC detectors.
Certified Values
Wavelength Values (nm)
Spectral Bandwidth 1 nm 4 nm 7 nm 10 nm
Peak Number 1 241.2 ± 0.2 241.1 ± 0.3 241.0 ± 0.3 241.1 ± 0.3
Peak Number 2 287.2 ± 0.2 287.8 ± 0.3 288.4 ± 0.3 287.5 ± 0.3
Peak Number 3 333.5 ± 0.2 333.2 ± 0.3 332.9 ± 0.3 332.8 ± 0.3
Peak Number 4 416.3 ± 0.2 417.2 ± 0.3 417.4 ± 0.3 417.5 ± 0.3
Peak Number 5 485.3 ± 0.2 485.3 ± 0.3 485.1 ± 0.3 484.2 ± 0.3
Peak Number 6 575.1 ± 0.2 575.3 ± 0.3 575.8 ± 0.3 576.2 ± 0.3
Peak Number 7 794.1 ± 0.2 794.4 ± 0.3 795.9 ± 0.3 797.2 ± 0.3
3 mL