A suitable supply of pure material was obtained.
This material is intended for the calibration of differential scanning calorimeters and
similar instruments.
Certified Values
Enthalpy of Fusion ............ 4.765 ± 0.012 kJ/mol
Melting Temperature .............. 327.47 ± 0.02 °C
A suitable supply of pure material was obtained.
This material is intended for the calibration of differential scanning calorimeters and
similar instruments.
Certified Values
Enthalpy of Fusion ............ 7.187 ± 0.011 kJ/mol
Melting Temperature .............. 231.92 ± 0.02 °C
A suitable supply of biphenyl was obtained and purified by repeated fractional
freezing from the melt.
This material is intended for the calibration of differential scanning calorimeters and
similar instruments.
Certified Values
Enthalpy of Fusion .............. 18.60 ± 0.11 kJ/mol
Melting Temperature ................ 68.93 ± 0.02 °C
A suitable supply of pure material was obtained.
This material is intended for the calibration of differential scanning calorimeters and
similar instruments.
Certified Values
Enthalpy of Fusion ............ 7.103 ± 0.034 kJ/mol
Melting Temperature .............. 419.53 ± 0.02 °C
A suitable supply of pure material was obtained.
This material is intended for the calibration of differential scanning calorimeters and
similar instruments.
Certified Values
Enthalpy of Fusion .......... 10.827 ± 0.052 kJ/mol
Melting Temperature .............. 660.33 ± 0.05 °C
A commercial supply of phenyl salicylate was obtained and purified by fractional
crystallisation.
This material is intended for the calibration of differential scanning calorimeters and
similar instruments.
Certified Values
Enthalpy of Fusion .............. 19.18 ± 0.08 kJ/mol
Melting Temperature ................ 41.79 ± 0.03 °C