A supply of n-nonane, of nominally 99 % purity, was obtained from a commercial
supplier.
This material is intended for use in validation work, or in quality control procedures,
for the determination of non-equilibrium flashpoint determined by the Abel closed cup
method as described in the Institute of Petroleum Standard IP170/95, and also
published as British Standard BS2000:Part 170: 1995.
Certified Value
Non-equilibrium Flashpoint ........................ 32.5 ± 0.5 °C
A supply of n-decane, of nominally 99 % purity, was obtained from a commercial
supplier.
This material is intended for use in validation work, or in quality control procedures,
for the determination of non-equilibrium flashpoint determined by the Abel closed cup
method as described in the Institute of Petroleum Standard IP170/95, and also
published as British Standard BS2000:Part 170: 1995.
Certified Value
Non-equilibrium Flashpoint ........................ 50.0 ± 0.9 °C