43
Descriptions of Additional Codes
Additional code Description (Applicable location)
01 Cooling plate (Cryo Module)
02 Parafn chamber (Embedding Module)
03 Left side warming chamber (Embedding Module)
04 Right side warming chamber (Embedding Module)
05 Left side hot plate (Embedding Module)
06 Right side hot plate (Embedding Module)
07 Parafn line (Embedding Module)
08 Dispensing solenoid valve (Embedding Module)
09 Dispenser (Embedding Module)
10 Forceps holder (Embedding Module)
11 Hot plate heater
21 Set value data
22 Operation status data
23 Dispensing log data
24 Error log data
25 Temperature record data
Buzzer Notication
This instrument noties the following events with a buzzer sound.
Type of notication Notication sound Condition for notication How to stop
Startup notication Short beep The display has started after the
power was turned on. (Approx. 15
seconds later)
Stops automatically.
Start notication Short beep The TEC 6-EM is operating. Stops automatically.
End notication Short beep Auto operation has ended. Stops automatically.
Error notication
(Caution)
One-minute continuous
beep
A caution-level error occurred Stops automatically or by
toughing the “
” (stop) button.
Error notication
(Warning)
Continuous beep A warning-level error occurred.
Touch the “
” (stop) button.
NOTE:
The buzzer notification cannot be disabled. The volume of the
buzzer is fixed.