42
Error
code
Overview
Displayed
message
Description
Instrument operation
when the error occurs
Solution
E103 Abnormally
high
temperature
Abnormally high
temperature
The temperature of each
location exceeds the
allowable value.
(The temperature is
higher than the E010
level.)
Operation aborts. urn off the power to the
instrument. When parafn in
the problem location starts
solidifying (below 60ºC, as
a guide), turn on the power
again. (The power should be
off for at least 10 seconds.)
If the problem persists,
turn off the power to the
instrument and contact
the Sakura instrument
distributor or representative.
(Check to see if any
equipment that generates
strong electromagnetic
waves is nearby.)
High temperature molten
parafn was added.
E901 Abnormal
nonvolatile
memory
Memory error
during data
update
Data could not be saved
because the nonvolatile
memory operated
abnormally.
Operation aborts. Operation aborts. Turn off
the power to the instrument
and contact the Sakura
instrument distributor or
representative.
E902
E903
E904