7-5. Standby operation
[Fig. 7-10] Standby operation setting screen
ItsetsY/Nofstandbyoperation.
Itsetsthetemperaturerangetobeappliedforstandbyoperation.
Itsetsthehumidityrangetobeappliedforstandbyoperation.
Itsetsthestandbytimetobeappliedwhentheindicateddatafor
temperatureorhumiditydoesnotenterintothestandbyoperation
range(TEMPPVorHUMIPV).
Itstandbysindefinitelyforentrytothestandbyoperationrangewhen
thestandbyoperationtimeissetin“00.00.”
Itdecideseitherof“Entire”and“MaintainSEG”forstandbyoperationmethod.
Entire:Thestandbyoperationisappliedtothesetentiresegmentin
[7-1Programpatternsetting]
MaintainSEG:Thestandbyoperationisappliedonlytotheset
maintainrangesegmentin[7-1Programpatternsetting]
●Itisascreentosetthetemperature&humidityStand-byRangeandStand-byTimeduringProgramoperation.
●Thesetstandbyoperationhereisappliedto[Fig.7-1Programpatternsetting].
●Definitionofstandbyoperation
-CoditionsforStand-byOperationEntry(“OR”condition):Incaseofnoentryofeitheroftemperatureorhumidityintothesetrangeforstandbyoperation
withinthesetsegmenttime
-ConditionsforStand-byOperationRelease(“AND”condition):Incaseofentryoftemperatureandhumidityintothesetrangeforstandbyoperation
-Thestandbytimehasindefinitevaluewhenthestandbytimeisnotset(Initialvalue).