6.Advanced parameters adjustment case
Order
Code
Items
Value in Case 1
Value in Case 2
1
F0
Temperature setting
25
2
F1
Temperature Difference
Values
-3
3
F2
Refrigeration return
difference
0.5
2.0
4
F3
Way of control
1
0
5
F4
Alarm for over high
water temperature
15
5
6
F5
Alarm for over low
water temperature
15
10
7
F6
Alarm for over high
room temperature
45
45
8
F7
Password
8
8
9
F8
The allowed highest water
temperature
31
30
10
F9
The allowed lowest water
temperature
25
5
11
F10
Alarm of sensor pulse
frequency
8
8
Case 1:Cooling water temperature is controlled by intelligent control mode. Requiring
water temperature to be between 25 to 31. The target water temperature is
3 lower than room temperature. When room temperature remains, the water
temperature fluctuation shall not exceed ±0.5.There will be an alarm when
water temperature is 15 lower or higher than target water temperature (e.g.
when room temperature is 30.0, cooling water temperature is between 27.5
to 26.5. If room temperature is up to 30.5, water temperature will be between
28.0 to 27.0).
Case 2:Cooling water temperature is controlled by constant temperature control mode.
Requiring water temperature to remain at 25, and the water temperature
fluctuation does not exceed ±2. The alarm of over high water temperature will
be on when water temperature is 5 higher than the set water temperature, and
the alarm of over low water temperature will be on when water temperature is 10
lower than the set water temperature. Users can easily adjust water temperature
between 5 and 30 through user setup.