OTHERS
PRESSURE
SENSOR
NORMAL
Error message
Troubled portion
Troubled phenomenon Check item Counteraction method
When there is pressure sensor
output of continuation 3 seconds
and less than 0.2V.
An abnormal power supply.
An abnormal power supply.
1. Check the volume rotary (body pressure) .
(CN903 red-black approx. 10k /red-white approx.
250 - 10k )
2. The connection confirmation of Printed board (Main) CN9
and Printed board (Massage pulse) CN901/CN903.
3. Check the main harness for disconnection or shorting.
4. Others.
1. Check the power supply of an power point.
2. Others.
1. Check the power supply of an power point.
2. Others.
Replace the volume rotary (body pressure).
Insert the connectors.
Replace the main harness.
Replace the Printed board (Main).
Replace the Printed board (Main).
Replace the Printed board (Main).
Up to seven abnormal indications are stored in a memory. In case of less than seven, since
there are not any abnormal indications, the STIFFNESS and TIMER indicators are disappeared
alternatively to indicate the state that abnormalities are not stored.
When the massaging rollers detach from the body during the Whole Body Sensor Automatic Course
or the Automatic Course, the body pressure sensor will work to stop the operation by pulling back
the massaging rollers for safety.
In such a case, begin the operation again from the start.
10 buttons flash on and off, and
the m
assage
stops