27
-
HEC-DR8700
Replacement Procedure
(7) Replacement of Power Switch
and Power Cord
1, Take off the arm rest (right) and the cover
(arm rest lower right). Photo-1,2
2, Remove two earthing setscrews around the
power supply box.
3, Remove two power supply box set screws
and take off the power supply box. Photo-11
4, Remove four case (power supply box upper)
set screws and take off the case (power
supply box upper). Photo-12
5, Remove each cord processing of the POWER
switch.
6, Take out the POWER switch forward while
pushing the top and bottom of the POWER
switch. Photo-13
7, Disconnect the connector of the power cord
and take off the power cord. Photo-13
8, Each reclining motor is fixed at two places
on both ends. Therefore, remove E rings or
R pins on both ends and pull out fixing pins.
Photo-10
(6) Replacement of Current Fuse
1, Open the fuse cover of the current fuse and
take out the current fuse. Photo-11
Current fuse
Fuse cover
Power supply box
set screws
Case (power supply box upper)
Case set screws
Power supply box
Photo-12
Photo-13
Photo-11
POWER switch
Connector (sensor A)
Connector (power cord)
Power cord
Reclining motor
(foot rest)
Reclining motor
(back rest)
Reclining motor (seat)
R rings & fixing pins
Photo-10
Reclining motor (back rest)
Reclining motor (foot rest)
E rings & fixing pins
Reclining motor (seat)