20
TROUBLESHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction Check/Remedy
The chamber is not cooled
at all
■ The voltage is too low. (In this case, call an electrician.)
■ The breaker is free.
■ The fuse is blown.
Sometimes the freezer is not cooled at all for about 5 minutes after
starting-up.
The cooling is poor
■ The ambient temperature is too high.
■ The door is not shut tightly.
■ The inner lid is not installed correctly.
■ The set temperature in the controller is not set properly.
■ The grille is blocked out.
■ The filter is clogged. Clean the filter when the filter check lamp
lights.
■ The freezer is in the direct sunlight.
■ There is any heating source near the freezer.
■ A rubber cap and insulation for the access port are not set
correctly.
■ You put too many unfrozen articles into the freezer compartment.
■ Too much frost is accumulated in the chamber.
Alarm test key cannot
actuate the alarm
■ The alarm test key does not function when the power switch of
the freezer is off.
■ When only the buzzer or only the alarm is actuated by the alarm
test key, the unactuated part is out of order, and must be replaced.
Noise
■ The freezer is not installed on the sturdy floor.
■ There is anything touching the frame.
■ The freezer is not leveled with the leveling feet.
■ It is the immediately after starting of refrigeration.
The freezer causes larger noise when the chamber temperature is
high because of large load. The noise gets smaller along with
chamber temperature.
Back-up test switch does
not operate normally (when
the back-up system is
provided) (AT type)
■ The liquid CO
2
cylinder is empty.
■ The valve of the liquid CO
2
cylinder is not opened.
■ The ambient temperature is too high. In this case, move the
liquid CO
2
cylinder to a cool location.
Inquire at liquid CO
2
suppliers about cylinder check, adjustment,
installation, or move.
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the
above table, contact Sanyo sales representative or agent.