VARIOUS SETTING (MENU/Log)
The operation of the hyper-terminal on the PC side
1. Transmission starts when a transfer, a capture of the text and a preservation file name are specified
and MENU/Start is selected.
2. The transmission is finished, with “Finished” indication.
Setup of communication: Log data receipt on the PC side (In case of Windows 2000 and XP)
1. Start a program-accessories-communication-hyper-terminal from the starting button. The registration of
the starting menu is to start in the following method, C : Start from ¥Program Files¥ Windows
NT¥hypertrm.exe.
2. Set up the following through the hyper-terminal screen.
New connection Name (Example) Sanyo
Setup of the connection Connection port COM1
Properties of COM1 Setup of the port
Bit/sec.: 9600 Data bit: 8 Parity: No Stop bit: 1 Flow control: Xon/Xoff
(When a log transmitting screen is opened, terms of communication on the MDF side are set on the
above condition automatically.)
3. The communication cable for the interface board MTR-480 of the option and 9 pin Dsub cross type for
RS232C is necessary for the data transmission to the PC. Refer to P. 40 for the details.
19

Other manuals for Sanyo MDF-C2156VAN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sanyo MDF-C2156VAN and is the answer not in the manual?

Sanyo MDF-C2156VAN Specifications

General IconGeneral
BrandSanyo
ModelMDF-C2156VAN
CategoryFreezer
LanguageEnglish

Related product manuals