EasyManuals Logo

Sanyo MDF-C2156VAN User Manual

Sanyo MDF-C2156VAN
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
REMOTE ALARM TERMINAL
The terminal of the remote alarm is installed at the panel door inside of the unit. The alarm is outputted
from this terminal. Contact capacity is DC 30 V, 2 A.
Contact output:
between COM. and N.O. between COM. and N.C.
At normal Open Close
At abnormal Close Open
Note:
The alarm is actuated when the power cord is disconnected from the outlet or the power switch is OFF.
REMOTE ALARM
MAX DC30 V 2 A
N.C. COM N.O.
BACK-UP SYSTEM
The holding time of chamber temperature due to the change in the ambient temperature.
MDF-C2156VAN
5
10
15
20
25
0 5 10 15 20 25 30 35
Temperature(℃)
Time (h)
• The above data use optional MDF-135N (Liquid N
2
gas cylinder).
25

Other manuals for Sanyo MDF-C2156VAN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sanyo MDF-C2156VAN and is the answer not in the manual?

Sanyo MDF-C2156VAN Specifications

General IconGeneral
BrandSanyo
ModelMDF-C2156VAN
CategoryFreezer
LanguageEnglish

Related product manuals