- 93 -
29
VARIOUS SETTING (MENU/Tools)
Set of door alarm delay time (Tools/Door Delay)
This product has the setup function of the door alarm delay time. (Factory setting is 2 minutes.) The
variable range of the door alarm delay time is 1-minute-15 minutes. . A setup example is the following.
Table 11 Door alarm delay time setup process (When door alarm delay time is changed from 2
minutes to 10 minutes.)
Description of operation Key operated Indication after operation
1 -----
The top screen displayed
2 Press the MENU key. MENU
The menu window is indicated. And the
“set” is highlighted.
3 Press the ENTER key.
The temp. setting window indicated.
4
Select Date Time item pressing Shift
keys.
A menu window is indicated, and the log of Date
Time
is highlighted.
5
Select Ring Back item pressing
MENU key. Then, pressing ENTER
key
The Date Time setting window indicated.
6
Select Door Delay item pressing
Shift keys.
Door Delay highlights.
7
The number input key is pressed in
order of 1ĺ0.
The numerical value of the Log Interval
(Log Interval) is changed at 15 to 10.
8 Press the MENU key. MENU
The menu window is indicated. And the
“OK” is highlighted.
9 Press the ENTER key. ENTER
Date time is memorized, and the top screen
window is indicated.
MENU
ENTER
ENTER
1
0

Related product manuals