11
Example
Determine an unknown percentage; store the weight on the balance/scale
as a reference percentage
Settings (changes in the factory settings required for this example):
Menu: Application program: Weighing in percent (2 1 5)
Menu: Reference percentage 100 % (3 3 5)
Step Key (or instruction) Display
1. Turn on the balance/scale e
2. Tare the balance/scale t 0.0 g
3. Display the reference v > 2 sec. ref100
percentage
4. Place the reference weight + 222.5 g
for 100% on the balance/scale
(here: 222.5 g)
5. Start application; if the print v + 100.0 %
format is set to include data
ID codes, Wxx% + 222.5 g
the following is printed
6. Place an unknown weight on the + 140.4 %
balance/scale (here: 322.5 g)
7. Display weight v + 222.5 g
8. Display percentage v + 100.0 %
9. Unload the balance/scale 0.0 %
10. Delete the reference percentage c
11. Repeat the procedure starting
from Step 6, if desired.