31
– After approx. 2 seconds the first linearization weight (user-defined weight)
is shown on the display. Select menu items 1-18-2 through 1-18-5 to enter the
linearization weights.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
– Press the ( key to stop the calibration/adjustment routine.
– Press the ) key to store the value for the first linearization weight;
the second linearization weight is prompted.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
– Press the ) key to store the value for the second linearization weight;
the third linearization weight is prompted.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).