Resolution  The resolution indicates the accuracy used to determine the reference weight. 
The default setting is “display resolution." The resolution is increased when “10-
fold" or “100-fold" is selected. “10-fold" increases the resolution of the net value 
by one step (display resolution x 10), “100-fold" increases it two steps (display 
resolution x 100). 
Setting: Applic./Applic.1/Count./RESOLUT menu item 3.9.
  Parameter for saving weight values  The weight on the platform is saved as a reference value as soon as the platform has 
stabilized. “Stability" is defi ned as the point at which the fl uctuation of a measured 
value lies within a defi ned tolerance. The narrower the tolerance, the more stable the 
platform is at “stability".
The “increased stability" setting has a lower tolerance so that the average piece 
weight saves is more accurate and the results more reproducible; however, the 
measurement time can take longer. 
Setting: Applic./Applic.1/Count./SAVE WT. menu item 3.11.
  Reference sample updating  You can defi ne whether or not the reference sample weight is updated automatically 
during weighing using this setting. The reference sample weight is updated 
automatically only when the following criteria are met:
1.   “Automatic" must be set for reference sample updating in the menu.
2.  The current piece count exceeds the original piece count by at least two.
3.   The current piece count cannot be more than double the size of the original 
piece count. This limitation does not apply to the fi rst update if the piece weight 
was entered via a barcode scanner or keypad.
4.  The current piece count is less than 1000.
5.   The internally calculated piece count (such as 17.24 pcs) differs by less than 
± 0.3 pcs from the nearest whole number (in this example: 17).
6.   The weighing platform is stable in accordance with the parameter defi ned for 
saving weights.
If automatic reference sample updating is selected in the menu and the piece count 
(pcs) is displayed, the Auto symbol is displayed below the bar graph. If the average 
piece weight has been updated since you began weighing, the text line shows the 
0pt symbol. During an updating operation, 0pt and the updated piece count are 
displayed briefl y in the measured value line. The new reference sample weight and 
reference sample quantity are saved. 
Setting: Applic./Applic.1/Count./REF.UPDT menu item 3.12.
  Counting with Two Weighing Platforms  You can use two weighing platforms simultaneously with the Counting application. 
When using two platforms, you can choose from the following operating modes:
–  Counting with two platforms of the same type
–  Counting with one reference platform and one weighing platform
Counting with two platforms of the same type
Use this mode to count different types of sample material with different weights. 
For example, count the lighter-weight pieces on one platform and the heavier pieces 
on another. You can defi ne one of the two scales as the reference scale. 
The reference scale is the fi rst scale active when you switch on the device, regardless 
of the setting for automatic initialization of the Counting application.
Counting with one reference platform and one weighing platform
In this operating mode, the reference platform is a high-resolution weighing 
platform with a relatively low maximum capacity. The other platform is used for 
weighing heavier samples, and has a high capacity with a relatively low resolution.
This allows you to both determine the reference sample weight with high resolution; 
i.e., very precisely, and to count large amounts of parts, without requiring an 
expensive high-resolution, high-capacity weighing platform.
56  Operating Instructions Combics Indicators
  Operation