–   Value saved automatically when the weighing platform is stable and the 
defi ned minimum load is exceeded.
  If the defi ned minimum load is not exceeded, you can save the item 
manually 
by pressing the O key. Regardless of these settings, the current value cannot 
be saved automatically unless the platform is unloaded before the next sample is 
placed on it. The weighing platform is considered to be unloaded when the load 
is less than 50% of the minimum load.
The number of items added to memory is shown in the number display.
 Press the c key to clear the totalizing memory. A printout is automatically 
generated.
 With two weighing platforms connected, you can add values from both platforms 
to the totalizing memory. The displayed result is accurately calculated in the active 
weight unit.
Example: When you add 1.243 g (determined on a weighing platform with three 
decimal places) to 1.4 g (determined on a platform with 1 decimal place), the display 
shows 2.643 kg.
  Preparation  t  Open the APPLIC. / APPLIC.3 / TOTALIZ menu.
  Available parameter settings   * = Factory setting
MIN.INIT Minimum load for initialization   3.6
  1 digit  1 digit  3.6.1*
  2 digit  2 digits  3.6.2
  5 digit  5 digits  3.6.3
  10 digT.  10 digits  3.6.4
  20 digT.  20 digits  3.6.5
  50 digT.  50 digits  3.6.6
  100 diG.  100 digits  3.6.7
  200 diG.  200 digits  3.6.8
  500 diG.  500 digits  3.6.9
  1000 dI.  1000 digits  3.6.10
AUTO.SAV  Autosave   3.16.
  OFF  Off  3.16.1*
  ON  On  3.16.2
PRT.SAV. Individual/Component printout when saved   3.17.
  OFF  Automatic printing off  3.17.1
  ON   Print the entire standard print confi guration 
every time with the O key is pressed 3.17.2*
VAL.FROM  Source of data for autosave   3.22.
  APPLIC.1  Application 1  3.22.1*
  APPLIC.2  Application 2  3.22.2
SAV.VAL. Save value    3.23.
  NET  Net  3.23.1*
  CALCUL.  Calculation  3.23.2
  NET+CAL  Net and Calculated  3.23.3
  Printout  You can confi gure whether a printout is generated automatically when a weight 
value is stored in the totalizing memory or manually by pressing the p key.
Setting: Applic./Applic.3/TOTALIZ menu item 3.17.
-   You can print manually by pressing the p key (single printout):  3.17.1 
-   Component log (single printout of an item): 3.17. 2
 The total data record is printed when you clear the totalizing memory (by pressing 
the c key).
86  Operating Instructions Combics Indicators
  Operation