EasyManuals Logo

Sartorius CP2202S User Manual

Sartorius CP2202S
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
15
Basic Weighing Function
Purpose
The basic weighing function is always
accessible and can be used alone or in
combination with an application program
(counting, weighing in percent, etc.).
Features
Taring the balance/scale
You can tare the balance/scale within the
entire weighing range.
Assigning IDs to weights (as needed)
Printing weights
Using Verified Balances/Scales as Legal
Measuring Instruments in the EU*:
The type-approval certificate for verification
applies only to non-automatic weighing
instruments; for operation with or without
auxiliary measuring devices, you must
comply with the regulations of your country
applicable to the place of installation of
your balance/scale.
Before using the balance/scale as a legal
measuring instrument, calibrate and adjust it
at the place of use using the built-in motor-
ized calibration and adjustment weight (span
adjustment device): see the “Calibration/
Adjustment" section in this chapter
$ The temperature range (°C) indicated on
the verification label may not be exceeded
during operation
Example:
BD BL 200
K +10 … +30 °C
* including the Signatories of the Agreement
on the European Economic Area

Other manuals for Sartorius CP2202S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius CP2202S and is the answer not in the manual?

Sartorius CP2202S Specifications

General IconGeneral
BrandSartorius
ModelCP2202S
CategoryScales
LanguageEnglish

Related product manuals