EasyManua.ls Logo

Sartorius CP2202S - Configuring the Balance;Scale

Sartorius CP2202S
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
External Calibration
Settings:
Calibration/adjustment mode: External calibration/adjustment (menu code 1 9 1)
The weight required for calibration/adjustment is defined in the factory settings
(see “Specifications").
Step Key (or instruction) Display
1. Turn on the balance/scale, e
0.0 g
if necessary.
2. Tare the balance/scale, t
0.0 g
if necessary.
3. Start calibration/adjustment q
+ 5000.0
Calibration weight is displayed CAL
without a unit symbol.
4. Apply the prompted calibration
5000.0
weight (in this example: CAL
5000 g).
After calibration, the weight
+ 5000.0 g
unit is shown. CAL
5. Remove the calibration weight.
0.0 g

Other manuals for Sartorius CP2202S

Related product manuals