Display Cause Solution
err 235 Connecting cable not Connect the cable correctly
on the CP225D connected correctly
Connection to junction Connect the equipment
on a different balance/scale correctly
The weight readout Unstable ambient Set up the balance/scale
changes constantly conditions in another area
A foreign object is caught Remove the foreign object
between the load plate and the
balance/scale frame
The weight readout The balance/scale has not Calibrate/adjust the
is obviously wrong been calibrated/adjusted balance/scale
The balance/scale was not Tare or zero the balance/scale
zeroed before weighing before weighing