Cleaning the Weighing Chamber and
Draft Shield
§ Open the draft shield cover and take out
the removable parts
§ Use a hand-held vacuum cleaner and
mini-hose to remove any powdered sample
material carefully.
§ Use blotting paper to remove any liquid sam-
ple material.
§ On models with a 3-sided draft shield,
pull the 3 draft shield walls upwards to
remove, if necessary.
Safety Inspection
If there is any indication that safe operation
of the balance/scale with the AC adapter
is no longer warranted:
§ Turn off the power and disconnect the
equipment from AC power immediately
> Lock the equipment in a secure place
to ensure that it cannot be used for the
time being
Safe operation of the balance/scale with the
AC adapter is no longer ensured when:
there is visible damage to the AC adapter.
the AC adapter no longer functions properly.
the AC adapter has been stored for a relatively
long period under unfavorable conditions.
Maintenance and repair work may be
performed only by service technicians who are
authorized by Sartorius and who
have access to the required maintenance man-
uals.
have attended the relevant service
training courses.
We recommend having the power supply
inspected by a certified electrician at
regular intervals, according to the checklist
given below:
Insulating resistance > 7 megaohms measured
with a constant voltage of at least 500 V at a
500 kohm load
Leakage current: < 0.05mA measured with a
properly calibrated multimeter
66