15
Basic Weighing Function
Purpose
The basic weighing function is always
accessible and can be used alone or in
combination with an application program
(counting, weighing in percent, etc.).
Features
Taring the balance/scale
You can tare the balance/scale within the
entire weighing range.
Assigning IDs to weights (as needed)
Printing weights
Using Verified Balances/Scales as Legal
Measuring Instruments in the EU*:
The type-approval certificate for verification
applies only to non-automatic weighing
instruments; for operation with or without
auxiliary measuring devices, you must
comply with the regulations of your country
applicable to the place of installation of
your balance/scale.
Before using the balance/scale as a legal
measuring instrument, calibrate and adjust it
at the place of use using the built-in motor-
ized calibration and adjustment weight (span
adjustment device): see the “Calibration/
Adjustment" section in this chapter
$ The temperature range (°C) indicated on
the verification label may not be exceeded
during operation
Example:
BD BL 200
K +10 … +30 °C
* including the Signatories of the Agreement
on the European Economic Area