Weighing in Percent
Long-term storage of the last reference
percentage “pRef” entered.
Toggling between percentage and
weight by pressing (F).
Function Keys
(F):
Begin calculation of percentage
> Current weight value stored as
reference weight “Wxx%” to be
loaded at initialization.
(CF):
End application program;
clear initialization data
Changing the reference percentage:
§ Press and hold (F) (2 sec.)
> Current reference percentage is
displayed.
$ Press (F) briefly to change the value;
press repeatedly until the desired
reference percentage is displayed.
Quantities to choose from: 1, 2, 5, 10,
20, 50, 100.
§ Store setting in long-term memory:
Press and hold (F).
Menu code: 2 1 5 *
Display symbol: %
Purpose
This application program allows you to
obtain weight readouts in percent which
are in proportion to a reference weight.
Features
The minimum load is equal to one digit,
defined according to the resolution of
the active weight unit.
Press and hold the (F) key (2 seconds)
to set the reference percentage
Storage parameter (rounding-off factor)
for storing the reference weight
to calculate the percentage can be
configured.
Configuration of decimal places dis-
played with a percentage.
Optional automatic output of the
reference weight “Wxx%” and reference
percentage to the data interface port
when the menu code 7 1 2 (print
application parameters) is set.
* =not available for CPA…-PCE models