110  Operating Instructions Cubis
®
 MCA
  Technical Data
15.7  Metrological Data
15.7.1  Models MCA225S | MCA225P | MCA125S | MCA125P
MCA225S  MCA225P MCA125S MCA125P
Unit Value Value Value Value
Scale interval (d) mg 0.01 0.01/0.02/ 
0.05
0.01 0.01 | 0.1
Maximum capacity (Max) g 220 60/120/220 120 60 | 120
Repeatability at 5% load
Standard deviation of the load values, 
tolerance
mg 0.015 0.015 0.015 0.015
Standard deviation of the load values, 
typical value
mg 0.0065 0.0065 0.0065 0.0065
Repeatability at approx. maximum capacity
Standard deviation of the load values, 
tolerance
mg 0.025 0.04 0.025 0.06
Standard deviation of the load values, 
typical value
mg 0.02 0.02 0.02 0.02
Linearity deviation
Tolerance mg 0.1 0.15 0.1 0.15
Typical value mg 0.065 0.1 0.065 0.1
Deviation when load is off-center, 
positions according to OIML R76
Test weight g 100 100 50 50
Tolerance mg 0.15 0.2 0.15 0.2
Typical value mg 0.1 0.1 0.1 0.1
Sensitivity drift between +10°C and +30°C ppK 1 1 1 1
Tare maximum capacity:  
Less than 100% ofmaximum capacity
Accuracy class according 
toDirective2014/31/EU
I I I I
Verification scale interval (e) according 
toDirective 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according 
toDirective2014/31/EU
mg 1 1 1 1
Minimum initial weighing according to USP 
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 8.2 8.2 8.2 8.2
Typical minimum initial weighing mg 13 13 13 13
Typical stabilization time s 2 2 2 2
Typical measurement time s 6 6 6 6