Talent and Gold Series
Model GE412, TE412, GE212, TE212, GE7101 TE6101,
TE412-L TE212-L TE6101-L
Weighing capacity g 410 210 7,100 6,100
Readability g 0.01 0.01 0.1 0.1
Tare range (subtractive) g 410 210 7,100 6,100
Repeatability <±g 0.01 0.01 0.1 0.1
Linearity <±g 0.02 0.02 0.2 0.2
Operating temperature range +10… +30°C (50° to 86°F)
Allowable ambient
operating temperature +5… +40°C (41°F to 104°F)
Sensitivity drift <±
within +10…+30 °C /K 0.5·10
–5
1 ·10
–5
0.5 ·10
–5
0.5 ·10
–5
Response time (average) s 2222
Adaptation to
ambient conditions By selection of 1 of 4 optimized filter levels
Display update (depends
on the filter level selected) s 0.2–0.8
External calibration weight g 200 (F2) 100 (M1) 5000 (F2) 5000 (F2)
(of at least accuracy class) lb* 0.4 0.2 – 10
Net weight, approx. kg 1.4 1.4 1.7 1.7
Pan size mm 116 d 116 d 174+143 174+143
Dimensions (B+T+H) mm 200+270+70
AC power source/
power requirements AC adapter 230 V or 115 V, +15% to –20%
Frequency 48–60 Hz
AC power source/
direct current V 10–20
Power consumption
(average) W 0.75
Hours of operation with
–
Mignon, AA alkaline
manganese battery,
approx. h 50
–
fully charged rechargeable
1000 mAh battery, approx. h 20
Hours of operation with
the YRB08Z rechargeable
battery pack h 25
* = only for TE models