29
Talent and Gold Series
Model GE4101, GE2101, GE811 TE601, TE12000,TE6100, TE4100,
TE4101, TE2101, …-L …-L …-L …-L
…-L …-L
Weighing capacity g 4,100 2,100 810 610 12,000 6,100 4,100
Readability g 0.1 0.1 0.1 0.1 1 1 1
Tare range (subtractive) g 4,100 2,100 810 610 12,000 6,100 4,100
Repeatability <±g 0.1 0.1 0.1 0.1 1 1 1
Linearity <±g 0.2 0.2 0.2 0.2 2 2 2
Operating temperature range +10…+30°C (50°F to 86°F)
Allowable ambient
operating temperature +5.… +40°C (41°F to 104°F)
Sensitivity drift
within +10…+30 °C <±/K 1·10
–5
2·10
–5
5·10
–5
5·10
–5
2.5·10
–5
5·10
–5
5·10
–5
Response time (average) s 2 1.5 1.5 1.5 1.5 1.5 1.5
Adaptation to
ambient conditions By selection of 1 of 4 optimized filter levels
Display update (depends
on the filter level selected) s 0.2–0.8
External calibration weight kg 2 (F2) 1 (M1) 0.5 (M1) 0.5 (M1)5 (M1) 5 (M1) 2 (M1)
(of at least accuracy class) lb* 4 2 1 1 10 10 4
Net weight, approx. kg 1.7
Pan size mm 174+143
Dimensions (B+T+H) mm 200+270+70
AC power source/,
power requirements AC adapter 230 V or 115 V, +15% to –20%
Frequency 48–60 Hz
AC power source/direct current V 10–20
Power consumption (average) W 0.75
Hours of operation with
– Mignon, AA alkaline
manganese battery, approx. h 50
– fully charged rechargeable
1000 mAh battery, approx. h 20
Hours of operation with
the YRB08Z rechargeable
battery pack h 25
* = only for TE models