44
Weighing in Percent
– Long-term storage of the last reference
percentage “pRef” entered.
– Toggling between percentage and
weight by pressing (F).
Function Keys
(F): 
Begin calculation of percentage
> Current weight value stored as 
reference weight “Wxx%” to be 
loaded at initialization.
(CF): 
End application program; 
clear initialization data
Changing the reference percentage:
§ Press and hold (F) (2 sec.) 
> Current reference percentage is 
displayed.
$ Press (F) briefly to change the value; 
press repeatedly until the desired 
reference percentage is displayed.
Quantities to choose from: 1, 2, 5, 10,
20, 50, 100.
§ Store setting in long-term memory: 
Press and hold (F).
Menu code: 2 1 5 *
Display symbol: %
Purpose
This application program allows you to
obtain weight readouts in percent which
are in proportion to a reference weight.
Features
– The minimum load is equal to one digit,
defined according to the resolution of
the active weight unit.
– Press and hold the (F) key (2 seconds) 
to set the reference percentage
– Storage parameter (rounding-off factor)
for storing the reference weight 
to calculate the percentage can be 
configured.
– Configuration of decimal places dis-
played with a percentage.
– Optional automatic output of the
reference weight “Wxx%” and reference
percentage to the data interface port
when the menu code 7 1 2 (print
application parameters) is set.
* =not available for CPA…-PCE models