18
Application Programs
Counting
Step Key (or instruction) Display
a. Select application program (Fn/Up) > 2 sec
b. Select “Counting” (Fn/Up) or (G/N/Down)
repeatedly
c. Press Print/Enter. (Print/Enter)
Display shows reference piece
selection menu.
Counting is blinking.
d. Automatically the display
will show the reference pieces. (Tare/Right)
Place container and press Tare
if required (display “tared”).
e. Select reference sample quantity: (Fn/Up) or (G/N/Down)
5, 10, 20, 50 or 100, and put repeatedly
the reference on the scale.
(quantity flashes)
f. Press Print/Enter . Place reference sample on scale
Now “O” flashes. Automatic optimization is
active. (Print/Enter)
g. Place more reference samples
on scale (The number is
x+2 ~ 2x times the reference
pieces (x) selected in step e.)
When the sample is still, the
reference sample weight is calculated.
The flashing “O” goes out.
h. Place uncounted parts/samples
on the scale
i. Toggle between reference piece weight, (G/N/Down)
total weight, and total pieces.
j. Unload the scale
k. Counting application: Clear (Zero/Left) > 2 sec