Example: Standard Drying with Fully Automatic Shutoff Mode
The moisture content of 2 g of corn starch is to be analyzed. The sample can scorch if overheated, but it is not particularly heat sensitive.
The analysis is to be ended automatically as soon as a constant weight is reached.
Settings (changes in the factory settings required for this example):
Program number: 1
Final temperature: 130°C
Start of analysis: With stability after “Enter” key is pressed
End of analysis: automatic (factory setting)
Part A: Configuring the Drying Parameters
Step Key (or Instruction) Display/Printout
1. Turn on the moisture analyzer e Sartorius logo is displayed
Self-test runs
2. Select the PROGRAM function y or x
3. Confirm the PROGRAM function u
4. Select P1 program u
5. Select Heating program u
6. Select Standard drying u
7. Enter the temperature: 130 u
3 ƒ x
u
5 ƒ y
u
27
PROGRAM
P1 105
o
C Auto.
S
0092
g
TARESETUP