Leveling the Moisture Analyzer
Purpose:
To compensate for unevenness at the place of installation
This is necessary especially for testing liquid samples that need to be at a uniform level
in the disposable sample pan
Always level the moisture analyzer again any time after it has been moved to a
different location.
§ Extend or retract the front and/or rear leveling feet as needed to adjust the moisture analyzer
Installing the Aluminum Panels (Optional)
! To prevent burns, allow the glass panels to cool sufficiently before removing them
! Do not handle the aluminum panels with oily or greasy fingers
!Do not scratch the aluminum panels; do not use abrasive or corrosive substances to clean
the aluminum panels
§ Remove the 6 screws and remove the panel retainer
§ Remove the glass panels
§ Position the aluminum panels in the retainer
§ Fasten the aluminum panels with the retainer and screws
Turning On the Analyzer; Opening and Closing the Sample Chamber
§ To turn on the analyzer: Press the e key
§ When opening or closing the sample chamber: Do not release the cover until it is in the fully
open or fully closed position
Setting the Language
See “Setting the Language” in the chapter entitled “Configuring the Moisture Analyzer”