External Calibration/ Adjustment with a Factory-Defined Weight
Externally calibrate and adjust the weighing system using a 30-g calibration weight
Step Key (or Instruction) Display/Printout
1. Select SETUP in the Function line x or y
2. Confirm SETUP u
3. Select isoTEST 2 ƒ y
4. Confirm isoTEST u
5. Confirm Weighing system settings u
6. Confirm Calibration/adjustment u
7. Unload and tare the weighing system u
8. Select the CAL. function y
9. Start calibration u
32
C Auto.
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
KAL.
WGH.SYS CAL./ADJ.Verzög. 2sec