EasyManua.ls Logo

Sartorius MA45 - Testing Communication and Heater; Testing SBI Communication; Testing the Heater Function

Sartorius MA45
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Testing SBI Communication
Step Key (or Instruction) Display/Printout
1. Prepare the test connector for the RS-232 port Connect TƒD (pin 2) with
(see “Pin Assignment Chart”) RƒD (pin 3)
2. Select “Hardware tests” see above
3. Select “Test interfaces” function u
and confirm
4. Confirm “SBI communication u
test” function
At the end of the test, the result
is displayed: Test error
or: Test OK (in this case, Test error)
(the test is repeated continuously)
5. Exit “Test SBI communication” 2 ƒ c
Testing the Heater
Step Key (or Instruction) Display/Printout
1. Select “Hardware tests” see the previous page
2. Select “Heater test” y
and confirm u
3. Start the heater test:
Select the START function u
After initialization, the current
temperature (in this case: 123°C)
and test time (in this case: 1.5 min)
are displayed
If the sample chamber is opened,
the test stops until the sample
chamber is closed again.
After a brief time interval, the device
reaches the nominal temperature.
4. Exit the heater test c
36
HARDWARE INTERFACE SBI
Result: Test Fehler
HARDWARE INTERFACE SBI
Result: Test error
25
END
HEATER TST 160
o
C Verzög. 2sec
S
o
C
STARTSETUP
123
ENDE
HEATER TST 160
o
C160
o
C 1.5min
S
o
C
STARTSETUP
160
ENDE
HEATER TST 160
o
C160
o
C 11.2min
S
o
C
STARTSETUP